Biochrom WPA Biowave DNAView larger

Lightwave II UV/VIS Spectrophotometer

80-3003-72

The Lightwave II diode array UV/Visible spectrophotometer combines ease of use with flexibility, incorporating Gifford optics with no moving parts and a Xenon source for long life high-energy performance.

The instrument includes a large wide view display and built-in software providing flash scan, fixed wavelength measurements, kinetics and concentration with comprehensive graphics and the ability to store up to 90 routines.

Samples may be measured in 10, 20 or 40 mm pathlength cells (glass, quartz or disposable). Instrument may be linked to a PC via a USB connection.

Features :

  • Split Beam (RBC) UV/Vis Spectrophotometer covering 190-1100nm (3/5nm bandwidth)
  • Compact, and excellent value for money compared to conventional systems.
  • Wavelength scanning, kinetics and concentration functionality with full graphics display
  • Flexible data export optional accessories include Export to PC (PVC) software,  integrated printer*, wireless Bluetooth® connectivity*, SD card accessory*, USB cable*

More details

Description
Specifications
Resources

Description

Compact and low maintainance
The Biochrom WPA Lightwave combines ease of use with flexibility, and the novel optics require no lid or moving parts  and use a Xenon source for high energy performance with longer lamp lifetime. Compact footprint saves valuable bench space. This instrument has a 2 year warranty and the Xenon lamp has a 3 year warranty. The Lightwave II+ is a higher resolution 3nm bandwidth version. 

Powerful and flexible
The instrument includes a large wide view display and in-built software providing flash scan, fixed wavelength measurements, kinetics and concentration with comprehensive graphics capability plus the ability to store up to 90 methods. On peak confirmation is also a feature of the flexible software. Concentration may be measured using either a factor, singlepoint calibration, multi-standard curves or alternatively there is a multi-wavelength mode where equations using absorbance values may be used for ratio calculations. Samples may be measured in 10, 20 or 40mm pathlength cells (glass, quartz or disposable).

Flexible data export options
Results may be printed to an optional integrated high quality graphical printer for a permanent record or exported via a USB connection to a suitable PC running Print Via Computer (PVC) software for advanced reporting and data storage or the optional wireless Bluetooth® accessory. Alternatively the optional SD card accessory allows laboratories to export to standard SD or SDHC memory cards. This will be particularly useful to laboratories wishing to clone methods and folders between instruments. Please note that only one optional accessory can be fitted to the instrument.

Ordering InformationOrder Code

Biochrom WPA Lightwave II with SD Card Accessory

80-3005-13

Biochrom WPA Lightwave II

80-3003-72

Biochrom WPA Lightwave II with Printer

80-3003-73

Biochrom WPA Lightwave II with Bluetooth®

80-3003-74

Biochrom WPA Lightwave II with SD Card Accessory

80-3005-13

Biochrom WPA Lightwave II+

80-3004-60

Biochrom WPA Lightwave II+ with Printer

80-3004-61

Biochrom WPA Lightwave II + with Bluetooth®

80-3004-62

Biochrom WPA Lightwave II+ with SD Card Accessory

80-3005-14

Upgrade and Spare parts
80-3004-07 Spare printer paper (20 rolls).
80-3003-84 Printer accessory.
80-3004-73 Print via computer software and cable.
80-3003-96 Bluetooth accessory
80-3005-00 SD Card accessory.

Specifications

SpecificationData
TypeUV/Vis 
Wavelength Range190 - 1100nm 
Wavelength Accuracy (+/-)2nm 
Wavelength Reproducibility (+/-)1nm 
Lamp SourceXenon (PTR) 
BeamSplit Beam 
Optical SystemMonochromator Holographic Grating 
Bandwidth5nm (3nm Lightwave II+) 
Detector1024 CCD array 
Photometric Range-0.300 to 2.500A, 0 to 199%T 
Photometric Accuracy+/- 0.003A at 0-0.5A at 546nm 
Photometric Reproducibility0.002A at 0 - 0.5A at 546nm 
Stray Light0.5%T at 220 & 340nm 
Stability0.01A/h at 340nm 
Measurement TypeFixed wavelength (Abs/%T), Abs ratio, concentration (standard curve), concentration (factor), multiwavelength, kinetics 
Cell CompatibilityCuvette, Microcell*, Long Pathlength Cell* 
Other Sample Containersn/a 
Cell HolderSingle Cell 
Accessory OptionsPrinter* 
Method StorageUp to 90 
ComplianceCE 
Data OutputPrint*, Store*, Export to PC 
DisplayGraphical LCD 
Data InterfaceUSB 
LanguagesEnglish 
Power Requirement90-250 V AC, 50/60Hz, 30VA 
Weight4.5Kg (9.9lb) 
Dimensions (WxDxH)10 x 26 x 39 cm (3.9 x 10.2 x 15.3 inches) 