MCA1601
The MagCore® HF16 is a simple, fast and cost-effective technique to automatically purify nucleic acids from a wide range of sample types for applications like gene expression analysis, clinical molecular diagnosis, human identity testing, forensics, and biomedical research :
This product is no longer in stock
Availability date:
The MagCore® HF16 System enables the parallel processing of up to 16 samples per run. With the pre-programmed protocols and magnetic bead based reagent cartridges, the system provides consistent and stable nucleic acid purification for every busy laboratory.
MagCore® reagent cartridges are pre-sealed incorporating all reagents needed for the purification process. No additional reagents are required and no handling is necessary. Pre-filled and sealed buffer cartridge and automated piercing step eliminate possible contamination or buffer spout. Patented Heating Well and Separation Well in the cartridge provide strong circulate force for Binding/Washing steps and give high purity in final eluate. The MagCore® beads are pre-allocated into each cartridge in Well 4. Mixing of the beads, reagents and sample is achieved by the liquid handling action of the instrument. Separation is achieved by magnetic attraction and liquid handling.
MagCore® Nucleic Acid Extraction Kits contain all reagents and consumables needed for 36, 72 and 96 nucleic acid isolation reactions. The consumables consist of reagent cartridges, individually packaged tip sets, sample tubes and elution tubes.
A wide range of ready-to-use kits is available for the MagCore system : >>>Kits for MagCore HF16
Model | HF16 |
---|---|
Capacity | Up to 16 samples in one run |
Protocols | Whole blood and body fluids (200 µl-2,400 µl) |
Viral Nucleic Acids (DNA/RNA) | |
Genomic DNA from tissue, plant... | |
RNA from whole blood, cultured cells... | |
Purification mode | Cellulose coated magnetic beads |
Sample volume | 200/400/1,200/2,400 µl |
Elution volume | Adjustable : 60/100/150/200 µl |
Yield | E.g.: 6 µg genomic DNA from 200 µl whole blood (capacity up to 25 µg) |
Purity | OD 1.8-2.0 A260/280 ratio |
Processing time | 30 to 70 minutes (depends on sample type and method) |
System components | 1. Pipetting unit : dispensing, transfering, X-Z two axis movement |
2. Electronic control : internal MP | |
3. UV light : power 5W, lifetime >6'000 hours | |
4. Heating block : from RT up to ~100°C | |
5. Display screen : 8 cm LCM screen with keypress panel | |
6. Accesories : T-rack, cartridge rack, empty cartridges, tips and tip holders, microcentrifuge tube, syringe O-ring and grease | |
Power supply | Voltage : AC100-240V Frequency : 50/60 Hz Power consumption >1.0 KVA |
Dimensions (W x D x H) | 55.0 x 66.0 x 68.0 cm |
Weight | 70 kg |
Brochure
Kit Selection Guide