6250000
The Coy Anaerobic Monitor (CAM-12) is the only analyzer that offers both O2 (PPM) and H2 (%) monitoring in one unit for the same price as most conventional Oxygen monitors.
Features :
This product is no longer in stock
Availability date:
The CAM-12 (and it’s predecessor, the Model 10) Analyzer have been successfully used by researchers to monitor levels of the oxygen and hydrogen in their anaerobic chambers for over 25 years. Your anaerobic chamber becomes a dynamic environment in which you can readily determine leaks, bad catalyst or a need for change in equipment layout or airlock procedure. The H2 monitor allows each lab to optimize their Gas Mix consumption rate saving time and operational cost.
Now labs who might otherwise have wanted separate, costly equipment to measure hydrogen can reap the additional benefits of knowing the levels of this volatile gas. Each analyzer is especially calibrated to your specific background gas mix. The security of knowing that an incorrect gas mixture can’t jeopardize your equipment or your safety is priceless.
Size of Footprint | 6.25” Wide x 6.25” Deep, 39.0 sq in (15.9 cm x 15.9 cm, 252.0 sq cm) |
Overall Dimensions | 6.25” Wide x 6.31” Deep x 3.35” High (15.9 cm x 16.03 cm x 8.52 cm) |
Weight | ~1.7 lbs (0.77 kg) (~3.0 lbs shipping weight) |
Power Requirements | ~18 watts 90 ~ 260VAC, 50 – 60 Hz |
Operating Environment | 15 to 42 ℃C Up to 90% Relative Humidity, Non-Condensing |
Full Scale Measurement Range | O2: 0 – 2000 ppm (readings will display higher levels) | |
H2: 0 – 10% (capable of higher values; however, unit should not be utilized in flammable mixtures) | ||
Resolution | O2: 1ppm from 0 -1500 ppm, and 5ppm from 1500 – 2000 ppm | |
H2: 0.1% from 0 – 4.0% | ||
Accuracy | O2: 20ppm up to 700ppm | |
H2: 1.0% up to 10.0% H2 | ||
Operating Temperature & Humidity Range | Ambient to 42℃ C and Non-Condensing | |
Zero Drift Due to Temperature | O2: < 1ppm / ℃C | The factory calibration temperature is 29 ℃C |
H2: < 0.1% / ℃C | ||
Output Display “Noise Level” | O2: +/- 1ppm | |
H2: +/- 0.1% | ||
Alarms – Oxygen | User-settable high alarm from 0 to 10,000ppm, in 5 ppm increments Flashing LED and message Audible tone can be silenced by the operator | |
Alarms – Hydrogen | User-settable between 1.5 and 4% in 0.1% increments Flashing LED and message Audible tone can be silenced by the operator (unless the Hydrogen is above 5%) | |
CO2 Compensation | User-settable from 1-20%* |
Enclosure | ABS plastic, 94V-0 |
Front, Rear, and Sensor Panel | Aluminum |
Sensor Handle | Anodized Aluminum |
Front Panel Overlay | Polycarbonate |
Rear Panel Overlay | Vinyl |
Display Face | Glass |
* If you are using a gas mix containing anything EXCEPT Hsub>2, Nsub>2 and/or <20% CO2, consult Coy Laboratory Products as this is outside of the standard CAM-12 range; however, a CAM-12S may be appropriate for your application. The CAM-12S is calibrated for use with gas mixes containing argon and/or >20% CO2 (the CO2 compensation function is potentially inaccurate or nonfunctional depending upon the gas mix).